  1. Does the dosage measurement add up values if an overrange condition (OVR+) is detected?

    • As of firmware 2.0 during overrange conditions the "end of range"-value is added to the total dosage and the color of the dose value changes to indicate that it is no longer precise due to overrange events.
    • in older Firmware all measurements during overrange conditions were not counted into the total dosage. In fact, the displayed dosage is too small and must be considered as wrong if the sensor was in overrange condition.

    Avoid overrange conditions during dosage measurements!

  2. How behaves the dosage measurement on if the node was temporarily disconnected?

    • During temporary disconnects of the node the dosage counter does not increase because the dosage increment is unknown.

Troubleshooting

Problem Solution
No connection possible Turn off both devices and restart.
If the problem persists, move display/collector closer to the sensors.
"OVR+" shown up in the display You are in overrange, that means, the irradiation is beyond the end of the measurement range.
=> Increase the distance between light source and sensor or attenuate the light by neutral density filters etc.
Note: any running dosage measurement is disturbed by overrange events, meaning the displayed dosage is lower as it should be.